Optional accessories – Leica Biosystems CM1950 User Manual

Page 62

Advertising
background image

62

Instructions for Use V 1.5 RevF - 08/2014

11. Optional Accessories

Adjusting the blade holder with anti-roll guide

You can adjust the height of the anti-roll guide

system using the knurled nut (1):

• If you turn the nut counterclockwise, the anti-

roll guide system moves toward the blade.

• If you turn the nut clockwise, the anti-roll guide

system moves away from the blade.

If the anti-roll guide system is in the wrong posi-

tion relative to the cutter, the following problems

will result:

Fig. I: The section rolls over the glass insert of the anti-roll guide system.
Error: Glass insert not high enough.
Remedy: Turn the knurled nut counterclockwise until the section is pushed between the

blade and anti-roll guide as shown in

Fig. III

.

Fig. II: Section tears and block hits the glass insert after sectioning.
Error: Anti-roll guide system is set too high.
Remedy: Turn the knurled nut clockwise until the section is pushed between the blade

and anti-roll guide as shown in Fig. III.

Fig. III: Correct position of the anti-roll guide to the cutter

I

II

III

1

Blade holder CE with

glass anti-roll guide

Generally, we recommend pre-adjusting the anti-roll guide system at a high section thickness

(e.g. 10 µm). Start from there and work your way down to the desired section thickness in small

increments, readjusting the anti-roll guide system at each increment using the knurled nut.

Fig. 70