11. Optional Accessories
Adjusting the blade holder with anti-roll guide
You can adjust the height of the anti-roll guide
system using the knurled nut (1):
• If you turn the nut counterclockwise, the anti-
roll guide system moves toward the blade.
• If you turn the nut clockwise, the anti-roll guide
system moves away from the blade.
If the anti-roll guide system is in the wrong posi-
tion relative to the cutter, the following problems
will result:
Fig. I: The section rolls over the glass insert of the anti-roll guide system.
Error: Glass insert not high enough.
Remedy: Turn the knurled nut counterclockwise until the section is pushed between the
blade and anti-roll guide as shown in
Fig. III
.
Fig. II: Section tears and block hits the glass insert after sectioning.
Error: Anti-roll guide system is set too high.
Remedy: Turn the knurled nut clockwise until the section is pushed between the blade
and anti-roll guide as shown in Fig. III.
Fig. III: Correct position of the anti-roll guide to the cutter
I
II
III
1
Blade holder CE with
glass anti-roll guide
Generally, we recommend pre-adjusting the anti-roll guide system at a high section thickness
(e.g. 10 µm). Start from there and work your way down to the desired section thickness in small
increments, readjusting the anti-roll guide system at each increment using the knurled nut.