Optional accessories – Leica Biosystems CM1950 User Manual
Page 58
58
Instructions for Use V 1.5 RevF - 08/2014
Assembling the anti-roll guide system
(for the blade holder CE)
1. Insert the glare-minimizing glass insert into the
interchangeable frame and tighten it evenly
using the knurled screw (10).
2. Insert the shaft (11) of the metal frame for
exchangeable glass inserts from above into
the bore of the swing arm (12) in such a way
that the pin rests in the notch.
3. Push the white plastic plate (14) from below
onto the shaft (11).
4. Screw the knurled nut (13) from below onto
the shaft (11).
13
14
11
6
10
10
• With blade ejector (1)
• With knife guard (2)
• integrated lateral shift and stable base
• Clearance angle adjustment (5) using size 4
Allen key (see Detail Fig. left on the blade holder)
– recommended angle between 2° and 5°.
• With anti-roll guide (4)
• Lever (6) for the lateral shift
• Lever (7) for clamping the blade
• Lever (8) for clamping the base (9) to the dove-
tail guide in the chamber
• Pressure plate (3) for section extraction
• When using low-profile blades, the straight-
edge
(7a,
Fig. 66
)
must be inserted.
Bladeholder CE with anti-roll guide (for low-profile, LP, and high-profile, HP)
1
4
5
2
3
11. Optional Accessories
Fig. 63a
4
6
7
8
9
12
The glass of the anti-roll guide can be
used from all 4 sides when replacement
is necessary (the glass stage plate can
be reordered).
Fig. 63
Fig. 64