Leica CM1950 – Cryostat

Reasons:

1. To prevent section waste from falling into the

opening.

2. To prevent cold from escaping from the chamber.

3. To prevent moisture from penetrating the

chamber.

6.

Installation

6.5.7 Inserting/changing the bacteria filter

6.5.8 Assembling the filter bag

Fig. 28

• Set the mark (A) of the extraction opening to

open (1) and pull it out. Plug the filter (3) into

the extraction hose connecting piece (3a) until

there is an audible click.

Now push the connected parts back into the

opening in the cryostat chamber (filter first) and

set it to the "closed" mark (2,

Fig. 28

).

1

3

3a

2

• To change the bacteria filter, follow the op-

posite procedure: press the filter to the right,

then pull it to the left and out of the tube.

• The filter must be changed approx. every

3 months (we recommend writing the date on

the filter using a marker).

Fig. 27

Fig. 29

A

E

L

The holder for the bacteria filter (optional) is vis-

ible in the front of the instrument.

• To insert the filter, hold it with one hand, press

on the right of the socket, then guide the filter

into the tube from the left.

The filter must be disposed of according

to valid laboratory specifications.

If completely defrosted, bacteria filters

and filter bags MUST be removed, as

melted coverslip mountant clogs the

extraction line. The bacteria filter ab-

sorbs the moisture during defrosting and

becomes unusable!

When not using the extraction, tightly

close the opening for the extraction

hose using the silicone stopper (E) in-

cluded in the scope of delivery.