Leica CM1950 – Cryostat

8.

Working with the Instrument

• Optimize the vacuum settings if necessary.

• Press the "VAC" button to enable the vacuum extractor. The LED in the

"VAC" button is lit while the extractor is on. Press the button again to

disable it.

• Use the knob to adjust the intensity of the vacuum.

A Optimal area for trimming and sectioning

B Optimal area for extraction from the chamber

• To clean the chamber, turn the knob to the red range.

A

B

The tension acting on the hose can be minimized

by turning the red ring (

Fig. 57

, top right) clock-

wise so that the suction nozzle presses against

the pressure plate (I,

Fig. 57

).

A

D

I

B

C

Fig. 57

K

• Check that the anti-roll guide is parallel and correctly adjusted. Readjust

if necessary (

see page 62

).

• Activate trimming mode.

• Select the trimming thickness.

• Switch on the VAC and select a low extraction level (between 1 and 2).

• Start trimming by manually moving the handwheel or press the

RUN/STOP and ENABLE buttons simultaneously to start motorized

sectioning.

Fig. 58

• Trimming:

Handwheel position 12 – 6 o'clock, valve open

Handwheel position 6 – 12 o'clock, valve closed

• Sectioning: Handwheel position 12 – 3 o'clock, valve open all the way

Handwheel position 3 – 6 o'clock, valve half open

Handwheel position 6 – 12 o'clock, valve closed

Ensure that the hose with the nozzle is

not installed against its "natural" curva-

ture on the pressure plate of the knife

holder.

For motorized sectioning, always begin at low speed for safety

reasons.