4.1 Installation
■■■■■■■■■■■■■■■■■■■■■■
94
■■■■■■■■
815 Robotic Soliprep
8 Select 103 under Rack position and enter the height 126 mm
under Work position Tower 1. Confirm with [OK].
9 Select the 2nd line (Special beakers 2) and click on [Edit].
10 Select 24 under Rack position and enter the height 100 mm
under Work position Tower 2. Confirm with [OK].
11 Select the 3rd line (Special beakers 3) and click on [Edit].
12 Select 104 under Rack position and enter the height 130 mm
under Work position Tower 1. Confirm with [OK].
13 The last position on the sample rack serves as Adjusting position.
Assign Special beakers 16 the rack position 105.
14 Save the Rack data by clicking on [OK].
15 Click [OK] on the tab.
16 Right-click on the instrument name 815_1 in the Devices window
and click on Initialize.
The Sample Processor is initialized. The settings that were made are
now valid.