background image

2.2 Setting lift positions

■■■■■■■■■■■■■■■■■■■■■■

36

■■■■■■■■

815 Robotic Soliprep

6 Click on the associated [Assign] button. Observe the illustration

above.

7 Switch back to the Move tab.

Lift position for the Polytron

1 In the manual control in the left-hand window, click on the Tower 2

item under 815_1 (815 Robotic …) and then select the Move tab.

2 Under Rack position, select the Target position = 1 and click on

[Start].

3 Move the lift to the Target position = Work position.

The Polytron must be at a suitable height for sample dispersion. Cor-
rect the lift position if necessary.

4 If the position has had to be corrected, switch to the Assign posi-

tion tab.

5 Under Lift position, select the Work position for = Tower, if this

is not yet selected.

6 Click on the associated [Assign] button.

7 Switch back to the Move tab.

Lift position for the washing station

These settings apply to tower 2.

1 Under Robotic arm position, select Target position = External

1 and click on [Start].

2 If the Polytron is not positioned over the washing station, correct the

position of the robotic arm using the arrow keys [Arrow left] and
[Arrow right] as necessary.

3 If the position has had to be corrected, switch to the Assign posi-

tion tab.

Advertising