Metrohm 815 Robotic Soliprep User Manual

Page 79

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Robotic Flexible Soliprep 2.815.2110

815 Robotic Soliprep

■■■■■■■■

71

2 Under Rack position, select the target position Special beaker 16

and click on [Start].

It is also possible to specify the target position as absolute rack posi-
tion. For the 6.2068.020 sample rack, the adjusting position no. is
105; for the 6.2068.030 sample rack the adjusting position no. is
115.

3 Enter 160 mm as target position under Lift position and click on

[Start]. Afterwards, move the lift further downward, one millimeter
at a time, until the robotic arm with the Luer adapter is located pre-
cisely above the positioning reticle.

The positioning reticle

The positioning reticle shows the directions of movement for the rack (R)
and the Swing Head (S). The tick marks stand for approx. 0.5° rotation
angle or swivelling angle deviation.

Determine the deviation of the adapter tip from the middle of the posi-
tioning reticle.

You can make the corresponding corrections in the tiamo configuration
afterwards.