3.3 Setting lift positions

■■■■■■■■■■■■■■■■■■■■■■

74

■■■■■■■■

815 Robotic Soliprep

This applies to the following positions:

Variable name

Position in mm

take needle

180

aspirate sample

150

take filter

167

put filter

127

waste tool

33

aspirate filtered

138

needle in vial

128

aspirate pressure

115

To find the method variables, proceed as follows:

1 Click on the Method symbol in the tiamo sidebar.

2 Click on Open... in the File menu.

3 Select 815 Robotic Flexible Soliprep - standard method from

the list of available methods and click on [Open].

4 Double-click on the START - Main track symbol.

5 Click on the Method variables tab.

The values of the variables can be changed in this table as necessary.

Opening the manual control

1 Click on the hand symbol in the tiamo sidebar.

2 In the left-hand window, click on the Tower 1 item under 815_1

(815 Robotic …) and then select the Move tab.