Metrohm 815 Robotic Soliprep User Manual
Page 82
3.3 Setting lift positions
■■■■■■■■■■■■■■■■■■■■■■
74
■■■■■■■■
815 Robotic Soliprep
This applies to the following positions:
Variable name
Position in mm
take needle
180
aspirate sample
150
take filter
167
put filter
127
waste tool
33
aspirate filtered
138
needle in vial
128
aspirate pressure
115
To find the method variables, proceed as follows:
1 Click on the Method symbol in the tiamo™ sidebar.
2 Click on Open... in the File menu.
3 Select 815 Robotic Flexible Soliprep - standard method from
the list of available methods and click on [Open].
4 Double-click on the START - Main track symbol.
5 Click on the Method variables tab.
The values of the variables can be changed in this table as necessary.
Opening the manual control
1 Click on the hand symbol in the tiamo™ sidebar.
2 In the left-hand window, click on the Tower 1 item under 815_1
(815 Robotic …) and then select the Move tab.