Metrohm 815 Robotic Soliprep User Manual
Page 130
4.4 Setting lift positions
■■■■■■■■■■■■■■■■■■■■■■
122
■■■■■■■■
815 Robotic Soliprep
Work height for sample beakers
1 Move to sample position 1. Under Rack position, enter the Target
position = 1 and click on [Start].
2 Fasten a syringe needle to the Luer adapter of the transfer robotic
arm.
3 Move the lift to the appropriate height for aspirating the dispersed
sample. Under Lift position, enter the Target position = 170 mm
and click on [Start].
If the needle is not in a suitable position afterwards, correct the value
for the aspirate sample method variable in the method. Repeat this
step.
4 Remove the needle from the Luer adapter.
Lift position for picking up needles and filters
1 Under Rack position, enter the Target position = 25 and click on
[Start].
2 Place a syringe needle in the established position in the insert.
3 Under Lift position, enter the Target position = 170 mm and
click on [Start]. The Luer adapter must grasp the needle and secure
it in place.