2 Robotic Titration Soliprep 2.815.1110

815 Robotic Soliprep

■■■■■■■■

35

Work height for sample beakers

The titration head should be fully equipped with electrode, dosing tips and
rinsing nozzles.

1 Move to sample position 1. Under Rack position, enter the Target

position = 1 and click on [Start].

2 Place a sample beaker on the established position.

3 Move the lift downward until the titration head is at a suitable height

for carrying out a titration. Under Lift position, click on the arrow
key [Arrow down] and also (if required) [Arrow up] in order to set
a suitable position.

4 Switch to the Assign position tab.

5 Under Lift position, select the Work position for = Tower, if this

is not yet selected.