3.3 Setting lift positions

■■■■■■■■■■■■■■■■■■■■■■

78

■■■■■■■■

815 Robotic Soliprep

Lift positions for special beakers

1 Under Rack position, enter the Target position = Special

beaker 1 and click on [Start].

2 Place a sample beaker on the established position. Lay the cover with

the hole on the sample beaker.

3 Fasten a membrane filter to the Luer adapter of the transfer robotic

arm.

4 Move the lift to the Target position = Work position. The filter

must rest on the cover of the sample beaker. Correct the lift position
if necessary.

5 If the position has had to be corrected, switch to the Assign posi-

tion tab.

6 Select the Work position for = Special beaker 1 in the Lift posi-

tion field.

7 Click on the associated [Assign] button.

8 Switch back to the Move tab.

9 In the manual control in the left-hand window, click on the Tower 2

item under 815_1 (815 Robotic …) and then select the Move tab.

10 Under Rack position, enter the Target position = Special

beaker 2 and click on [Start].

11 Place a suitable sample beaker on the established position and fill it

with a sponge or a crushed absorbent paper towel. This is to be used
to dab off the moist Polytron aggregate.