Metrohm 815 Robotic Soliprep User Manual

Page 134

Advertising
background image

4.4 Setting lift positions

■■■■■■■■■■■■■■■■■■■■■■

126

■■■■■■■■

815 Robotic Soliprep

8 Click on the associated [Assign] button.

9 If the position of the robotic arm has been changed, select External

1 under Robotic arm position.

10 Click on the associated [Assign] button.

11 Switch back to the Move tab.

12 Under Robotic arm position, select External 2 as Target posi-

tion and click on [Start].

The Luer adapter should now be located in the opening of the
deflector and touch it on the right.

13 Correct the position of the robotic arm as necessary using the arrow

keys [Arrow left] and [Arrow right].

14 If the position has had to be corrected, switch to the Assign posi-

tion tab.

15 Under Robotic arm position, select the External 2 and click on

the associated [Assign] button.

16 Switch back to the Move tab.

Advertising