3 Robotic Flexible Soliprep 2.815.2110

815 Robotic Soliprep

■■■■■■■■

79

12 Move the lift to the Target position = Work position. The Poly-

tron aggregate must sufficiently come into contact with the absorb-
ent material. Correct the lift position if necessary.

13 If the position has had to be corrected, switch to the Assign posi-

tion tab.

14 Select the Work position for = Special beaker 2 in the Lift posi-

tion field.

15 Click on the associated [Assign] button.

16 Switch back to the Move tab.

Lift positions for stripping needles and filters

1 In the manual control in the left-hand window, click on the Tower 1

item under 815_1 (815 Robotic …) and then select the Move tab.

2 Fasten a syringe needle or a membrane filter to the Luer adapter of

the transfer robotic arm.

3 Under Robotic arm position, select the Target position = Exter-

nal 1 and click on [Start].

4 Move the lift downward until the thick sleeve of the Luer adapter is

approx. 1 cm underneath the deflector. Under Lift position, click on
the arrow key [Arrow down] and also (if required) [Arrow up] in
order to set a suitable position.

The Luer adapter must be positioned at a distance of approx. 2 cm
from the deflector. Its thick sleeve must be positioned approx. 1 cm