Metrohm 815 Robotic Soliprep User Manual

Page 133

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Robotic Filtration Soliprep 2.815.3110

815 Robotic Soliprep

■■■■■■■■

125

Lift positions for stripping needles and filters

1 In the manual control in the left-hand window, click on the Tower 1

item under 815_1 (815 Robotic …) and then select the Move tab.

2 Fasten a syringe needle or a membrane filter to the Luer adapter of

the transfer robotic arm.

3 Under Robotic arm position, select the Target position = Exter-

nal 1 and click on [Start].

4 Move the lift downward until the thick sleeve of the Luer adapter is

approx. 1 cm underneath the deflector. Under Lift position, click on
the arrow key [Arrow down] and also (if required) [Arrow up] in
order to set a suitable position.

The Luer adapter must be positioned at a distance of approx. 2 cm
from the deflector. Its thick sleeve must be positioned approx. 1 cm
beneath the deflector in order for the adapter to move into the
opening of the deflector.

5 To correct the position of the Luer adapter, correct the lift position

with the arrow keys [Arrow down] and [Arrow up] first. Then cor-
rect the position of the robotic arm using the arrow keys [Arrow
left]
and [Arrow right].

6 Switch to the Assign position tab.

7 Select the Work position for = External 1 in the Lift position

field.