4 Robotic Filtration Soliprep 2.815.3110
815 Robotic Soliprep
■■■■■■■■
113
1
2
3
4
5
1 Screw an adsorber tube filled with cotton to the Vent connector on
the dosing unit.
2 Equip the 10 mL dosing unit with a filling tube (Port 2 on the under-
side of the dosing unit) and screw it onto the 6.1608.023 amber
glass bottle (with GL 45 thread). The bottle should be filled with the
solvent (water or an organic solvent) which is also used for sample
dispersion.
3 Place the bottle with the dosing unit into the 6.2061.010 bottle
holder.
4 Connect the transfer tubing to tower 1 with the end that is still free
to Port 1 on the dosing unit.
5 Attach the Dosino (dosing drive) onto the dosing unit.
6 Equip the connection cable of the Dosino with an anti-interference
adapter made of ferrite (T.2400.102) (see figure) and connect it to
the MSB 1 connector on the rear of the Sample Processor.