2 Robotic Titration Soliprep 2.815.1110

815 Robotic Soliprep

■■■■■■■■

25

5

5 Press on the pressure clamp and clamp tightly with the locking lever.

Tighten the clamping screw until the pump tubing cannot shift posi-
tion. The flow rate of the pump can be adjusted later with the
clamping screw while the pump is running.

6 Connect the connection cable of the pump to tower 2.

Plug the threaded plug of the connection cable into the connec-
tion socket Ext. Pump 2 on the rear of the tower.
Correct alignment of the 3 contact pins must be observed.

Tighten the knurled screw at the front end of the plug by hand in
clockwise direction. This will secure the plug.

Peristaltic pump for aspirating the titration vessel
Proceed as follows for connecting and attaching the tubing:

For illustrations, see preceding section.

1 Cut one piece of 6.1826.100 pump tubing down to a length of

approx. 17 cm and attach a 6.1820.050 tubing olive at both ends.