Metrohm 815 Robotic Soliprep User Manual

Page 34

Advertising
background image

2.1 Installation

■■■■■■■■■■■■■■■■■■■■■■

26

■■■■■■■■

815 Robotic Soliprep

2 Fasten a piece of 6.1812.000 PTFE tubing leading to a waste con-

tainer to one of the tubing olives. Attach a piece of 6.1812.000 PTFE
tubing approx. 1 m long to the other olive connector. Connect the
6.1805.220 aspiration tubing to the other end of the tubing with the
aid of the 6.1820.030 M8 tubing adapter and a 6.1828.000 connec-
tion nipple.

6.1805.220

6.1820.030

6.1828.000

6.1812.000

3 Release the locking lever by rotating the clamping screw and open

the pressure clamp.

4 Loop the pump tubing around the rotor and fasten with the tubing

clamps. The rotor turns in clockwise direction. The inlet tubing must
therefore be attached below, the outlet tube above.

5 Press on the pressure clamp and clamp tightly with the locking lever.

Tighten the clamping screw until the pump tubing cannot shift posi-
tion. The flow rate of the pump can be adjusted later with the
clamping screw while the pump is running.

6 Connect the connection cable of the pump to tower 1.

Advertising