3 Robotic Flexible Soliprep 2.815.2110

815 Robotic Soliprep

■■■■■■■■

77

5 Correct the lift position and the respective aspirate filtered method

variable if necessary.

6 Move the lift upward and remove the needle.

7 Fasten a membrane filter to the Luer adapter of the transfer robotic

arm.

8 Move the lift to the Target position = 127 mm. The filter must rest

on the sample vial.

9 Correct the lift position and the respective put filter method variable

if necessary.

10 Move the lift upward and remove the filter.

Lift position for vials

1 Under Rack position, enter the Target position = 83 and click on

[Start].

2 Place a sealed sample vial in the established position in the insert.

3 Fasten a syringe needle to the Luer adapter of the transfer robotic

arm.

4 Move the lift to the Target position = 128 mm.

The needle must perforate the vial and be at a suitable height for
releasing the filtered sample.

5 Correct the lift position and the respective needle in vial method

variable if necessary.

6 Move the lift to the Target position = 115 mm.

The needle tip must be located above the sample liquid in the vial.

7 Correct the lift position and the respective aspirate pressure

method variable if necessary.

8 Move the lift upward and remove the needle.