Metrohm 815 Robotic Soliprep User Manual
Page 85
■■■■■■■■■■■■■■■■■■■■■■
3 Robotic Flexible Soliprep 2.815.2110
815 Robotic Soliprep
■■■■■■■■
77
5 Correct the lift position and the respective aspirate filtered method
variable if necessary.
6 Move the lift upward and remove the needle.
7 Fasten a membrane filter to the Luer adapter of the transfer robotic
arm.
8 Move the lift to the Target position = 127 mm. The filter must rest
on the sample vial.
9 Correct the lift position and the respective put filter method variable
if necessary.
10 Move the lift upward and remove the filter.
Lift position for vials
1 Under Rack position, enter the Target position = 83 and click on
[Start].
2 Place a sealed sample vial in the established position in the insert.
3 Fasten a syringe needle to the Luer adapter of the transfer robotic
arm.
4 Move the lift to the Target position = 128 mm.
The needle must perforate the vial and be at a suitable height for
releasing the filtered sample.
5 Correct the lift position and the respective needle in vial method
variable if necessary.
6 Move the lift to the Target position = 115 mm.
The needle tip must be located above the sample liquid in the vial.
7 Correct the lift position and the respective aspirate pressure
method variable if necessary.
8 Move the lift upward and remove the needle.