4.1 Installation
■■■■■■■■■■■■■■■■■■■■■■
90
■■■■■■■■
815 Robotic Soliprep
4 Click on the Tower 2 tab and then on Configuration.
Enter the following settings:
■
Swing offset = 0.0°
■
Max. swing range = 115.5°
■
Swing radius = 110 mm
■
Rotation offset = 0.0°
■
Swing direction = +
5 Finish the settings with [OK].
The instrument must be reinitialized in order to make the settings
effective.
6 Click [OK] on the tab.
7 Right-click on the instrument name 815_1 in the Devices window
and click on Initialize.
The Sample Processor is initialized. The settings of the Swing Heads
and robotic arms are now valid.
Advertising