3.1 Installation

■■■■■■■■■■■■■■■■■■■■■■

48

■■■■■■■■

815 Robotic Soliprep

8 Select 113 under Rack position and enter the height 126 mm

under Work position Tower 1. Confirm with [OK].

9 Select the 2nd line (Work position Tower 2) and click on [Edit].

10 Select 114 under Rack position and enter the height 100 mm

under Work position Tower 2. Confirm with [OK].

11 The last position on the sample rack serves as Adjusting position.

Assign Special beakers 16 the rack position 115.

12 Save the Rack data by clicking on [OK].

13 Click [OK] on the tab.

14 Right-click on the instrument name 815_1 in the Devices window

and click on Initialize.

The Sample Processor is initialized. The settings that were made are