4.1 Installation
■■■■■■■■■■■■■■■■■■■■■■
108
■■■■■■■■
815 Robotic Soliprep
5
5 Press on the pressure clamp and clamp tightly with the locking lever.
Tighten the clamping screw until the pump tubing cannot shift posi-
tion. The flow rate of the pump can be adjusted later with the
clamping screw while the pump is running.
6 Connect the connection cable of the pump to tower 2.
■
Plug the threaded plug of the connection cable into the connec-
tion socket Ext. Pump 2 on the rear of the tower.
Correct alignment of the 3 contact pins must be observed.
■
Tighten the knurled screw at the front end of the plug by hand in
clockwise direction. This will secure the plug.