4.4 Setting lift positions

■■■■■■■■■■■■■■■■■■■■■■

124

■■■■■■■■

815 Robotic Soliprep

5 If the position has had to be corrected, switch to the Assign posi-

tion tab.

6 Select the Work position for = Special beaker 1 in the Lift posi-

tion field.

7 Click on the associated [Assign] button.

8 Switch back to the Move tab.

9 Move the lift upward and remove the membrane filter.

10 In the manual control in the left-hand window, click on the Tower 2

item under 815_1 (815 Robotic …) and then select the Move tab.

11 Under Rack position, enter the Target position = Special

beaker 2 and click on [Start].

12 Place a sample beaker on the established position.

13 Move the lift to the Target position = Work position. The Luer

adapter must be at a suitable position for rinsing the transfer tubing.
Correct the lift position if necessary.

14 If the position has had to be corrected, switch to the Assign posi-

tion tab.

15 Select the Work position for = Special beaker 2 in the Lift posi-

tion field.

16 Click on the associated [Assign] button.

17 Switch back to the Move tab.