4.4 Setting lift positions
■■■■■■■■■■■■■■■■■■■■■■
124
■■■■■■■■
815 Robotic Soliprep
5 If the position has had to be corrected, switch to the Assign posi-
tion tab.
6 Select the Work position for = Special beaker 1 in the Lift posi-
tion field.
7 Click on the associated [Assign] button.
8 Switch back to the Move tab.
9 Move the lift upward and remove the membrane filter.
10 In the manual control in the left-hand window, click on the Tower 2
item under 815_1 (815 Robotic …) and then select the Move tab.
11 Under Rack position, enter the Target position = Special
beaker 2 and click on [Start].
12 Place a sample beaker on the established position.
13 Move the lift to the Target position = Work position. The Luer
adapter must be at a suitable position for rinsing the transfer tubing.
Correct the lift position if necessary.
14 If the position has had to be corrected, switch to the Assign posi-
tion tab.
15 Select the Work position for = Special beaker 2 in the Lift posi-
tion field.
16 Click on the associated [Assign] button.
17 Switch back to the Move tab.