3.1 Installation

■■■■■■■■■■■■■■■■■■■■■■

44

■■■■■■■■

815 Robotic Soliprep

4 Click on the Tower 2 tab and then on Configuration.

Enter the following settings:

Swing offset = 0.0°

Max. swing range = 115.5°

Swing radius = 110 mm

Rotation offset = 0.0°

Swing direction = +

5 Finish the settings with [OK].

The instrument must be reinitialized in order to make the settings
effective.

6 Click [OK] on the tab.

7 Right-click on the instrument name 815_1 in the Devices window

and click on Initialize.

The Sample Processor is initialized. The settings of the Swing Heads
and robotic arms are now valid.