4 calibration function, Press the x button – KEYENCE IL Series User Manual

Page 107

Advertising
background image

4-4 Calibration Function

4-53

IL-E

4

Setting V

ar

iou

s F
unct

ions

[

SEt3]

8

Insert workpiece B for which the measured value is
known, and using the
S / T buttons, set the "CALC
you wish to display (3)" for that workpiece B.

The current internal measurement value is displayed on the main
display (upper level). [SEt3] and the "CALC you wish to display
(3)" are alternately displayed on the sub display (lower level).
IL-S025/IL-030/IL-S065/IL-065/IL-100

IL-300/IL-600

IL-2000

[SET] button

9

Press the [SET] button. (SET3 confirmation operation)

After [SEt3] blinks several times on the sub display (lower
level), [End] is displayed.
This completes the calibration.

X

button

10

Press the X button.

The main screen is restored.



LASER

BANK

0

1

2

3

HI

GO

LO

R.V. ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

HOLD

CALC

CHECK



CALC value

Setting you wish to display

Head connected
to main unit

Head connected to
expansion unit

Workpiece B

Item

Setting range

Default value

set3

-99.999 to 99.999

10.000

Item

Setting range

Default value

set3

-999.99 to 999.99

100.00

Item

Setting range

Default value

set3

-9999.9 to 9999.9

1000.0

'P&

LASER

BANK

0

1

2

3

HI

GO

LO

R.V. ANALOG

HI SHIFT

ZERO SHIFT

TIMING

LO

HOLD CALC CHECK

Main screen

Advertising