Setting the shift target value, Setting the shift target value -18 – KEYENCE IL Series User Manual

Page 50

Advertising
background image

3-18

IL-E

3

Basic Oper

ations

3-5

Zero Shift Function (Shifting the Internal Measurement Value (R.V.))

The internal measurement value (R.V.) is shifted (offset) to an arbitrary shift target value.
The judgment value (P.V.) is shifted (offset) as well.
On a main unit using the calculation function, the calculated value (CALC value) will be
shifted (offset) to the shift target value instead of R.V.
The following two methods can be used.
• Press the [ZERO SHIFT] button (within 1 second).
• Set the external input (zero shift input) to ON for 20 ms or more.*

* When the zero shift input is set for the external input 1 (pink wire), the zero shift is

enabled when the input is triggered..

"11. External Input" (page 4-30)

1

Press the W / X button several times on the main screen. To display the
shift target value on the sub display (lower level).

"Sub Display (Lower Level)" (page 3-4)

2

Press S / T button to set the shift target value.

IL-S025/IL-030/IL-S065/IL-065/IL-100

IL-300/IL-600

IL-2000

3

Press W / X button to return the sub display to the original screen.

Setting the Shift Target Value

Item

Setting range

Default value

Shift target value

-99.999 to 99.999

0.000

Item

Setting range

Default value

Shift target value

-999.99 to 999.99

0.00

Item

Setting range

Default value

Shift target value

-9999.9 to 9999.9

0.0