Analog output accuracy, Analog output accuracy -29 – KEYENCE IL Series User Manual

Page 83

Advertising
background image

4-2 Basic Settings and Advanced Settings

4-29

IL-E

4

Setting V

ar

iou

s F
unct

ions

„

Setting range for analog output scaling

If free range or bank is selected for the analog output scaling state, the upper limit and
lower limit can be set within the range shown below.
IL-S025/IL-030/IL-S065/IL-065/IL-100

IL-300/IL-600

IL-2000

The analog output can be selected from ± 5 V, 1-5 V, 0-5 V or 4-20 mA.

*1 The averaging is as follows after 512 times.

512 times: Sampling rate x 2
1024 times: Sampling rate x 4
2048 times: Sampling rate x 8
4096 times: Sampling rate x 16

Item

Setting range

Default value

Upper limit

-99.999 to 99.999

10.000

Lower limit

-99.999 to 99.999

-10.000

Item

Setting range

Default value

Upper limit

-999.99 to 999.99

100.00

Lower limit

-999.99 to 999.99

-100.00

Item

Setting range

Default value

Upper limit

-9999.9 to 9999.9

1000.0

Lower limit

-9999.9 to 9999.9

-1000.0

Analog Output Accuracy

Voltage output

Current output

Output range

± 5 V
(Full scale 10 V)

4 - 20 mA
(Full scale 16 mA)

Output resistance

100

Ω

-

Maximum load resistance

-

350

Ω

Repeatability

± 1 mV

± 1.5 µA

Accuracy in respect to
display

± 0.05% of F.S.

± 0.25% of F.S.

Temperature
characteristics

± 0.005% of F.S./°C

± 0.01% of F.S./°C

Refresh rate

*1

Same as sensor head sampling rate

Advertising