4-2 Basic Settings and Advanced Settings

4-33

IL-E

4

Setting V

ar

iou

s F
unct

ions

z

When Peak hold, Bottom hold or Peak-to-peak hold is set
"7. Timing Input" (page 4-23) setting: "Level"

"7. Timing Input" (page 4-23) setting: "Edge"

z

When Auto peak hold or Auto bottom hold is set

„

Minimum Input Time

ON time: 2 ms
OFF time: 2 ms

Max. 2 ms

Max. 2 ms

Min. 2 ms

Min. 2 ms

ON

OFF

Max. 2 ms

No sampling

No sampling

Sampling period

Previous result

Judgment value
(P.V.) judgment
output

Timing Input

Judgment value
(P.V.) judgment
output

Max. 2 ms

Max. 2 ms

Min. 2 ms

Min. 2 ms

ON

OFF

Timing Input

Sampling period 1

Sampling period 1 result

Sampling period 2 result

Previous result

Sampling period 2

Sampling period 3

Max. 2 ms

Max. 2 ms

Max. 2 ms

Min. 2 ms

Min. 2 ms

ON

OFF

Timing Input

Sampling does not start

Sampling starts

Advertising