4-2 Basic Settings and Advanced Settings

4-27

IL-E

4

Setting V

ar

iou

s F
unct

ions

The analog output scaling for the judgment value (P.V.) is set.

If the analog output is set to no output, this screen is skipped.

“3-1 Operation When the Power is Turned on for the First Time” (page 3-2)
“3-3 Initial Reset (Initialize)” (page 3-8)

„

Relation of voltage and analog output in default state (dEFLt)

The upper and lower limits for the different selections of analog outputs are as follows.

10. Analog Output Scaling

#P)

Analog output

Description

Default

value

Default setting

Scaling is not performed.

c

Free range

Scaling is performed. The analog output range can be
changed by specifying the upper limit and lower limit of the
judgment value (P.V.).

Bank

A different scaling is provided for each bank. The analog
output range can be changed by designating an upper
limit and lower limit for each bank.

Example 1: Lower limit < Upper limit

(Analog output 1 to 5 V)

Example 2: Lower limit > Upper limit

(Analog output 1 to 5 V)

Reference

If Upper limit = Lower limit is set, the analog output is output as below.
Judgment value (P.V.)

≤ Upper limit: Analog output lower limit (When 1 to 5 V → 1 V)

Judgment value (P.V.)

> Upper limit: Analog output upper limit (When 1 to 5 V → 5 V)

Lower limit Upper limit

5

1

Output upper
limit

Output lower
limit

Judgment
value
(P.V.)

(V)

(V)

5

1

Lower limit Upper limit

Judgment
value
(P.V.)

Output upper
limit

Output lower
limit

Model

Upper limit

Lower limit

IL-S025

5.000

- 5.000

IL-030

5.000

- 5.000

IL-S065/IL-065

10.000

- 10.000

IL-100

20.000

- 20.000

IL-300

140.00

- 140.00

IL-600

400.00

- 400.00

IL-2000

1000.0

-1000.0

Analog setting

Output upper limit

Output lower limit

OFF

Not output

0-5v

5 V

0 V

-5-5v

5 V

- 5 V

1-5v

5 V

1 V

AMPr

20 mA

4 mA