1 setting operations, Setting operations, 1 setting operations -2 – KEYENCE IL Series User Manual

Page 56: Setting operations -2, Functions available on the main screen

Advertising
background image

4-2

IL-E

4

Setting V

ar

iou

s F
unct

ions

4-1

Setting Operations

This section explains functions operable on the main screen and functions operable after
the display changes to each setting screen.

Functions Available on the Main Screen

Setting Operations

DIN rail mount type (IL-1000/IL-1050)

Panel mount type (IL-1500/IL-1550)

Buttons used

Main screen

Press W or X button.

Switching display on the sub display (lower level) (page 3-4)

The internal measurement value (R.V.), the analog output
value, the HIGH side setting value, the LOW side setting value
or Calculation value (CALC value) is displayed and the setting
can be changed.

3-4 Setting the Tolerance Setting Value (page 3-10)

HIGH side setting value and LOW side setting value are set. The judgment is
made among HIGH/GO/LOW, and the value is displayed and output.

3-5 Zero Shift Function (Shifting the Internal Measurement

Value (R.V.)) (page 3-18)

The internal measurement value (R.V.) can be shifted (offset)
to an arbitrary shift target value.

While pressing down [MODE], press S or T button.

3-6 Bank Function (Registering Multiple Tolerance Setting

Values) (page 3-20)

HIGH side setting value, LOW side setting value, shift target
value, and analog output scaling upper / lower limit value can
be saved in up to four banks and switched.

Press [MODE] and S buttons for approx. 2 seconds.

or

Press [MODE] and T buttons for approx. 2 seconds.

3-7 Key Lock Function (page 3-21)

This function prevents unwanted button operations during measurement.



LASER

BANK

0

1

2

3

HI

GO

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT



HOLD

CALC

CHECK