Calibration Function

When there is a difference between the internal measurement value (R.V.) and the actual
dimension of the object, the value can be calibrated.
The zero point (offset) and tilt are both corrected.

When calculation function is disabled

Setting method (Calibrating R.V.)

Point

• If the following conditions are not satisfied, the screen below is

displayed when the setting is complete. Make the setting again.

• Workpiece is not within measurement range.
• The setting was switched during the calibration.

Main screen

1

Press [MODE] button and X button for approx. 2
seconds on the main screen at the same time.

[30.CLb] is displayed on the main display (upper level).

2

Press S / T button to set [USr] and press X button.

Press [MODE] +
X

button

2 seconds or more
at the same time

Calibration

R.V. you wish to display (2) [

5'V]

R.V. before calibration (2)

R.V. you wish to display (1) [

5'V]

R.V. before calibration (1)

Offset value A

Measurement value

Displayed value

'TT

LASER

BANK

0

1

2

3

HI

GO

LO

R.V. ANALOG

HI SHIFT

ZERO SHIFT

TIMING

LO

5'6

HOLD CALC CHECK

0.5 (R.V. you wish to display (2) ) - (R.V. you wish to display (1))

(R.V. before calibration (2)) - (R.V. before calibration (1))

2



LASER

BANK

0

1

2

3

HI

GO

LO

R.V. ANALOG

HI SHIFT

ZERO SHIFT

TIMING

LO



HOLD CALC CHECK

Setting

value

Description

Default

value

defLt

Standard

c

Usr

User setting

*'#&

LASER

BANK

0

1

2

3

HI

GO

LO

R.V. ANALOG

HI SHIFT

ZERO SHIFT

TIMING

LO

%#.KD

HOLD CALC CHECK

Calibration setting