Display color, Display color -41, Or "18. display color – KEYENCE IL Series User Manual

Page 95

Advertising
background image

4-2 Basic Settings and Advanced Settings

4-41

IL-E

4

Setting V

ar

iou

s F
unct

ions

Near measurement center

IL-S025 : 25 mm ± 0.25 mm

IL-030 : 30 mm ± 0.25 mm

IL-S065/IL-065 : 65 mm ± 0.5 mm

IL-100 : 100 mm ± 1 mm

IL-300 : 300 mm ± 7 mm

IL-600 : 600 mm ± 20 mm

IL-2000: 2000 mm ± 50 mm

*1 When the analog output of the main unit is set to OFF, or the expansion unit is operating, the

indicator turns on if the following distances are detected.
IL-S025: 25 ± 5 mm
IL-030: 30 ± 5 mm
IL-S065/IL-065: 65 ± 10 mm
IL-100: 100 ± 20 mm
IL-300: 300 ± 140 mm
IL-600: 600 ± 400 mm
IL-2000: 2000 mm ± 1000 mm

*2 If GO is set to red with "18. Display color", the green LED and red LED lighting methods will

be reversed.

*3 The position of the display light (CENTER/A.RANGE) of the IL-S025/IL-S065/IL-2000 is

different from that of above.

The LED color of the judgment indicator is set. The display color for the main display (upper
part) can be set at the same time only for the panel mount type amplifier (IL-1500/IL-1550).

“3-4 Setting the Tolerance Setting Value” (page 3-10)

CENTER

A.RANGE

LASER

IL-030

(1) Reference distance indicator

*3

(2) Analog range indicator

*3

(3) Laser radiation emission indicator

18. Display Color

%Q.

Display color

DIN rail mount type

(IL-1000/1050)

Panel mount type

(IL-1500/IL-1550)

Default

value

Judgment indicator Judgment indicator

Main display

GO Green

HI: Red
GO: Green
LO: Red

HI: Red
GO: Green
LO: Red

GO: Green,
otherwise: Red

c

GO Red

HI: Green
GO: Red
LO: Green

HI: Green
GO: Red
LO: Green

GO: Red, otherwise:
Green

Always Red

HI: Red
GO: Red
LO: Red

HI: Red
GO: Red
LO: Red

Always Red display