Calibrating the voltage probes – Fluke Biomedical 190M User Manual

Page 109

Advertising
background image

Maintaining the Test Tool

Calibrating the Voltage Probes

7

99

Calibrating the Voltage Probes

To meet full user specifications, you need to adjust the
voltage probes for optimal response. The calibration
consists of a high frequency adjustment and a dc
calibration for 10:1 probes and 100:1 probes. The probe
calibration matches the probe to the input channel.

This example shows how to calibrate the 10:1 voltage
probes:

1

A

Display the input A key labels.

2

F3

Open the

PROBE ON A

menu.

If the correct probe type is already
selected (yellow shaded) you can
continue at step 5.

3

ENTER

Select Probe Type: Voltage, and
Attenuation: 10:1.

4

F3

Open the

PROBE ON A

menu again.

5

F1

Select PROBE CAL…

Probe Cal

Probe Cal Ref.

Figure 50. Adjusting Voltage Probes