Automatic connect-and-view trigger, Edge trigger, Isolated external trigger (190m-2) – Fluke Biomedical 190M User Manual

Page 122

Advertising
background image

190M Series Medical ScopeMeter
Users Manual

112

Max. Delay ................................................... 60 s at 5 s/div

Automatic Connect-and-View Trigger

Source ............................................................... A, B, C, D

EXT

(190M-2)

Slope ............................................ Positive, Negative, Dual

Edge Trigger

Screen Update ........... Free Run, On Trigger, Single Shot

Source ....................................... A, B, C, D, EXT (190M-2)

Slope ........................................... Positive, Negative, Dual

Trigger Level Control Range ...........................

±

4 divisions

Trigger Sensitivity
DC to 5 MHz at >5 mV/div ........................... 0.5 divisions
DC to 5 MHz at 2 mV/div and 5 mV/div ........... 1 division
200 MHz (Fluke Biomedical190M-2) ................ 1 division
250

MHz

.........................................................

2

divisions

Isolated External Trigger (190M-2)

Bandwidth ................................................................. 3 kHz

Modes ...................................................... Automatic, Edge

Trigger Levels (DC to 3 kHz) ...................... 120 mV, 1.2 V