N-cycle triggering – Fluke Biomedical 190M User Manual

Page 73

Advertising
background image

Triggering on Waveforms

Triggering on Edges

4

63

N-Cycle Triggering

N-Cycle triggering enables you to create a stable picture
of for example n-cycle burst waveforms.

Each next trigger is generated after the waveform has
crossed the trigger level N times in the direction that
complies with the selected trigger slope.

To select N-Cycle triggering, continue from step 3 (page
61) again:

4

ENTER

Select On Trigger or Single Shot,
jump to Trigger Filter.

5

ENTER

Select a Trigger Filter or set it Off.

6

ENTER

Set NCycle to On

Observe that the key labels at the bottom of the screen
have been changed to allow further selection of specific
N-Cycle trigger settings:

7

Set the number of cycles N

8

Adjust the trigger level

Traces with N-Cycle triggering (N=2) and without N-Cycle
triggering are shown in Figure 31.

Figure 31. N-Cycle triggering