190M Series Medical ScopeMeter
Users Manual

124

Environmental

Environmental .......................... MIL-PRF-28800F, Class 2

Temperature
Operating:
battery installed ..................... 0 to 40

°

C (32 to 104

°

F)

no

battery

installed

................ 0 to 50

°

C (32 to 122

°

F)

Storage ............................. -20 to +60

°

C (-4 to +140

°

F)

Humidity (Maximum Relative)
Operating:
0

°

C to 10

°

C (32

°

F to 50

°

F) .............. noncondensing

10

°

C to 30

°

C (50

°

F to 86

°

F) .............. 95 % (

±

5 %)

30

°

C to 40

°

C (86

°

F to 104

°

F) ..............75 % (

±

5 %)

40

°

C to 50

°

C (104

°

F to 122

°

F) ............45 % (

±

5 %)

Storage:
-20 to +60

°

C (-4 to +140

°

F) ............... noncondensing

Altitude
Operating:
CATIII 600V, CATII 1000V ............. 3 km (10,000 feet)
CATIV 600V, CATIII 1000V .............. 2 km (6,600 feet)
Storage ............................................ 12 km (40,000 feet)

Vibration (Sinusoidal) ............................................ max 3 g

Vibration (Random) ........................................... 0.03 g

2

/Hz

Shock ................................................................... max 30 g

Electromagnetic Compatibility (EMC)
Emissions and immunity ....... EN/IEC61326-1 (2005-12)

Enclosure Protection .......................... IP51, ref: IEC60529

Certifications

Conforms to .....

(CE),

(CSA),

(N10140)

Safety

Designed for 1000 V

CAT III

, 600 V

CAT IV

,

Pollution Degree 2, per:

EN/IEC 61010-1:2001 Pollution Degree 2
(According to CE mark)

IEC61010-031:2002+A1:2008

ANSI/UL 61010-1:2004 {ed. 2.0}

CAN/CSA C22.2 No. 61010-1-04 (including approval)