Triggering on video frames, Triggering on video lines – Fluke Biomedical 190M User Manual

Page 76

Advertising
background image

190M Series Medical ScopeMeter
Users Manual

66

Triggering on Video Frames

Use FIELD 1 or FIELD 2 to trigger either on the first half of
the frame (odd) or on the second half of the frame
(even).To trigger on the second half of the frame, do the
following:

7

F1

Choose FIELD 2.

The signal part of the even field is displayed on the
screen.

Triggering on Video Lines

Use ALL

LINES

to trigger on all line synchronization

pulses (horizontal synchronization).

7

F2

Choose

ALL LINES.

The signal of one line is displayed on the screen. The
screen is updated with the signal of the next line
immediately after the test tool triggers on the horizontal
synchronization pulse.

To view a specific video line in more detail, you can select
the line number. For example, to measure on video line
123, continue from step 6 as follows:

7

F3

Enable video line selection.

8

Select number 123.

The signal of line 123 is displayed on the screen. Observe
that the status line now also shows the selected line
number. The screen is continuously updated with the
signal of line 123.