Fluke Biomedical 190M User Manual

Page 15

Advertising
background image

Safety Information: Read First

5

XW

Warning

To avoid electrical shock or fire, use only
power cords and plugs that meet local safety
regulations with the supplied BC190/808
Universal Power Adapter.

Note:

To accommodate connection to various line
power sockets, the BC190/808 Universal Power
Adapter is equipped with a male plug that must
be connected to a line cord appropriate for local
use. Since the adapter is isolated, the line cord
does not need to be equipped with a terminal for
connection to protective ground. Line power
cords with a protective grounding terminal are
more commonly available. It is OK to use
grounded line power cords, even though the
ground terminal is not required.

XW

Warning

To avoid electrical shock or fire if a test tool
input is connected to more than 42 V peak,
30 V RMS or 60 V DC:

Use only insulated voltage probes, test leads

and adapters supplied with the test tool, or
indicated by Fluke Biomedical as suitable for
the 190M Series Medical ScopeMeters.

Before use, inspect voltage probes, test leads

and accessories for mechanical damage and
replace when damaged.

Remove all probes, test leads and

accessories that are not in use.

Always connect the power adapter first to the

ac outlet before connecting it to the test tool.

Do not touch voltages more than 42 V peak ,

30 V RMS or 60 V DC.

Do not connect the ground spring (Figure 1,

item d) to voltages more than 42 V peak, 30 V
RMS, or 60 V DC with respect to earth
ground.

Do not apply more than the rated voltage,

between the terminals or between each
terminal and earth ground.