Using the Scope and Meter

Acquiring Waveforms

1

25

Acquiring Waveforms

Setting the Acquisition Speed and Waveform
Memory Depth

To set the acquisition speed, do the following:

1

SCOPE

Display the

SCOPE

key labels.

2

F4

Open the

WAVEFORM OPTIONS

menu.

3

ENTER

Select Acquisition:
Fast –
for fast trace update rate;
shortest record length, decreased
zoom rate, no readings possible.
Full – maximum waveform detail;
10,000 samples per trace record
length, maximum zoom rate,
lower trace update rate.
Normal – optimal trace update
rate and zoom range combination

4

F4

Exit the menu

See also Table 2 in Chapter 8.


Selecting AC-Coupling

After a reset, the test tool is dc-coupled so that ac and dc
voltages appear on the screen.
Use ac-coupling when you wish to observe a small ac
signal that rides on a dc signal. To select ac-coupling, do
the following:

1

A

Display the

INPUT A

key labels.

2

F2

Highlight

AC

.

Observe that the bottom left of the screen displays the
ac-coupling icon:

.

You can define how Auto Set affects this setting, see
Chapter 6 “Changing the Auto Set Options.”