116

Full Scale Reading ................ 999 counts (3-digit display)

Accuracy (logic or pulse) ....................

±

(0.5 % +2 counts)

Pulse Width (PULSE)

Resolution (with

GLITCH

off) ......................... 1/100 division

Full Scale Reading .......................................... 999 counts

Accuracy
1 Hz to full bandwidth ......................

±

(0.5 % +2 counts)

Vpwm

Purpose ................ to measure on pulse-width modulated

signals, like motor drive inverter outputs

Principle ........ readings show the effective voltage based

on the average value of samples over a
whole number of periods of the fundamental
frequency

Accuracy ............................. as Vrms for sinewave signals

V/Hz

Purpose .................... to show the measured Vpwm value

(see Vpwm) divided by the fundamental frequency

on Variable-Speed AC Motor drives.

Accuracy ..................................................... %Vrms + %Hz

Note

AC motors are designed for use with a rotating
magnetic field of constant strength. This strength
depends on the applied voltage (Vpwm) divided
by the fundamental frequency of the appied
voltage (Hz). The nominal Volt and Hz values are
shown on the motor type plate.

Power (A and B, C and D)

Power Factor ........................ ratio between Watts and VA
Range

..........................................................

0.00

to

1.00

Watt ................................... RMS reading of multiplication:

corresponding samples of input A or C (volts)

and Input B or D (amperes)

Full

Scale

Reading

.......................................

999

counts

VA ................................................................. Vrms x Arms
Full

Scale

Reading

.......................................

999

counts

VA Reactive (VAR) ..........................................

((VA)

2

-W

2

)

Full Scale Reading ....................................... 999 counts