Meter measurements for model 190m-2, Meter input (banana jacks), Meter functions – Fluke Biomedical 190M User Manual

Page 128: General, Ohms ((), Continuity (cont), Diode

Advertising
background image

190M Series Medical ScopeMeter
Users Manual

118

Meter Measurements for Model 190M-2

The accuracy of all measurements are within

±

(% of reading + number of counts) from 18

°

C to 28

°

C.

Add 0.1 x (specific accuracy) for each

°

C below 18

°

C or

above 28

°

C.

Meter Input (Banana Jacks)

Input Coupling .............................................................. DC

Frequency Response ......................... DC to 3 kHz (-3 dB)

Input Impedance ................ 1 M

Ω

(±1 %)//14 pF (±1.5 pF)

Max. Input Voltage: ............................. 1000 V CAT III

600

V

CAT

IV

(For detailed specifications, see “Safety”)

Meter Functions

Ranging ........................................................ Auto, Manual

Modes ...................................................... Normal, Relative

General

DC Common Mode Rejection Ratio (CMRR)......... >100 dB
AC Common Mode Rejection at 50, 60, or 400 Hz ....... >60 dB

Ohms (

Ω

)

Ranges .................................. 500.0

Ω

, 5.000 k

Ω

, 50.00 k

Ω

,

500.0

k

Ω

, 5.000 M

Ω

, 30.00 M

Ω

Full Scale Reading
500

Ω

to 5 M

Ω

............................................... 5000 counts

30

M

Ω

............................................................ 3000 counts

Accuracy .................................................

±

(0.6 % +5 counts)

Measurement Current .................... 0.5 mA to 50 nA,

±

20 %

decreases with increasing ranges

Open Circuit Voltage ..................................................... <4 V

Continuity (CONT)

Beep ............................................................. <50

Ω

(

±

30

Ω

)

Measurement Current .................................. 0.5 mA,

±

20 %

Detection of shorts of .................................................

1 ms

Diode

Maximum Voltage Reading ....................................... 2.8 V

Open Circuit Voltage .................................................. <4 V

Accuracy .................................................

±

(2 % +5 counts)

Measurement Current ............................... 0.5 mA,

±

20 %