Using vertical cursors on a waveform – Fluke Biomedical 190M User Manual

Page 64

Advertising
background image

190M Series Medical ScopeMeter
Users Manual

54

Using Vertical Cursors on a Waveform

To use the cursors for a time measurement (T, 1/T), for a
mVs-mAs-mWs measurement, or for an RMS
measurement of the trace section between the cursors, do
the following:

1

CURSOR

From scope mode, display the
cursor key labels.

2

F1

Press to highlight

3

F3

Choose, for example, time
measurement:

T

.

4

F4

Choose the trace that you want
the markers to be placed on:

A

,

B

,

C

,

D

or

M

(Mathematics).

5

F2

Highlight the left cursor.

6

Move the left cursor to the
desired position on the waveform.

7

F2

Highlight the right cursor.

Figure 26. Time Measurement with Cursors

8

Move the right cursor to the
desired position on the waveform.

The screen shows the time difference between the cursors
and the voltage difference between the two markers. See
Figure 26.

9

F4

Select OFF to turn off the cursors.