Using the Scope and Meter

Using Average, Persistence and Glitch Capture

1

21

Using Average, Persistence and Glitch
Capture

Using Average for Smoothing Waveforms

To smooth the waveform, do the following:

1

SCOPE

Display the

SCOPE

key labels.

2

F4

Open the

WAVEFORM OPTIONS

menu.

3

Jump to Average:

4

ENTER

Select On... to open the

AVERAGE

menu.

5

ENTER

Select Average factor: Average
64
. This averages the outcomes
of 64 acquisitions.

6

ENTER

Select Average: Normal (normal
average) or Smart (smart
average, see below)

You can use the average functions to suppress random or
uncorrelated noise in the waveform without loss of
bandwidth. Waveform samples with and without
smoothing are shown in Figure 9.

Smart average

In the normal average mode occasional deviations in a
waveform just distort the averaged wave shape, and do
not show up on screen clearly. When a signal really
changes, for instance when you probe around, it takes
quite some time before the new wave shape is stable.
With smart averaging you can quickly probe around, and
incidental waveform changes like a line flyback in video
show up on screen instantly.