190M Series Medical ScopeMeter
Users Manual

114

Automatic Scope Measurements

The accuracy of all readings is within

±

(% of reading +

number of counts) from 18

°

C to 28

°

C. Add 0.1x (specific

accuracy) for each

°

C below 18

°

C or above 28

°

C. For

voltage measurements with 10:1 probe, add probe
accuracy, see section ‘10:1 Probe’ on page 127. At least
1.5 waveform periods must be visible on the screen.

General

Inputs .............................................................. A, B, C and D

DC Common Mode Rejection Ratio (CMRR)......... >100 dB
AC Common Mode Rejection at 50, 60, or 400 Hz ....... >60 dB

DC Voltage (VDC)

Maximum Voltage
with 10:1 probe ...................................................... 1000 V
direct (1:1) .................................................................300 V

Maximum Resolution
with

10:1

probe

..........................................................

1 mV

direct (1:1) .............................................................. 100 µV

Full Scale Reading ............................................. 999 counts

Accuracy at 5 s to 10

μ

s/div

2

mV/div

..........................................

±

(1.5 % + 10 counts)

5 mV/div to 100 V/div .........................

±

(1.5 % + 5 counts)

Normal Mode AC Rejection at 50 or 60 Hz ............. >60 dB

AC Voltage (VAC)

Maximum Voltage
with

10:1

probe

....................................................

1000

V

direct

(1:1)

.............................................................

300

V

Maximum Resolution
with 10:1 probe ....................................................... 1 mV
direct (1:1) ........................................................... 100 µV

Full Scale Reading .......................................... 999 counts

Accuracy
DC

coupled:

DC

to

60

Hz

.................................

±

(1.5 % +10 counts)

AC coupled, low frequencies:
50

Hz

direct

(1:1)

........................

±

(1.5 % + 10 counts)

60

Hz

direct

(1:1)

........................

±

(1.9 % + 10 counts)

With the 10:1 probe the low frequency roll off point will
be lowered to 2 Hz, which improves the AC accuracy
for low frequencies. When possible use DC coupling
for maximum accuracy.

AC or DC coupled, high frequencies:
60 Hz to 20 kHz ..........................

±

(2.5 % + 15 counts)

20 kHz to 1 MHz ............................

±

(5 % + 20 counts)

1 MHz to 25 MHz .........................

±

(10 % + 20 counts)

For higher frequencies the test tool’s frequency roll off
starts affecting accuracy.
Normal Mode DC Rejection .............................. >50 dB