Fluke Biomedical 190M User Manual

Page 110

Advertising
background image

190M Series Medical ScopeMeter
Users Manual

100

A message appears asking you whether to start the 10:1
probe calibration.

6

F4

Start the probe calibration.

A message appears telling you how to connect the probe.
Connect the red 10:1 voltage probe to input A and to the
probe calibration reference signal as shown in Figure 50.

7

Adjust the trimmer screw in the
probe housing until a pure square
wave is displayed.

For instructions to access the
trimmer screw in the probe housing
see the probe instruction sheet.

8

F4

Continue with DC calibration.
Automatic DC calibration is only
possible for 10:1 voltage probes.

The test tool automatically calibrates itself to the probe.
During calibration you should not touch the probe. A
message indicates when the DC calibration has completed
successfully.

9

F4

Return.

Repeat the procedure for the blue 10:1 voltage probe on
input B, the gray 10:1 voltage probe on input C and the
green 10:1 voltage probe on input D.

Note

When using 100:1 voltage probes, choose 100:1
attenuation to perform an adjustment.