Triggering on External Waveforms
(Model 190M-2)

Use external triggering when you want to display
waveforms on inputs A and B while triggering on a third
signal. You can choose external triggering with automatic
triggering or with edge triggering.

1

Supply a signal to the red and black 4-mm
banana jack inputs.

In this example you continue from the Trigger on Edges
example. To choose the external signal as trigger source,
continue as follows:

2

TRIGGER

Display the

TRIGGER

(On Edges)

key labels.

3

F1

Select Ext (external) edge trigger.

Observe that the key labels at the bottom of the screen
have been adapted to allow selection of two different
external trigger levels: 0.12 V and 1.2 V:

4

F3

Select 1.2V under the Ext LEVEL
label.

From this point the trigger level is fixed and is compatible
with logic signals.