Using cursors on spectrum measurements – Fluke Biomedical 190M User Manual

Page 65

Advertising
background image

Using Replay, Zoom and Cursors

Making Cursor Measurements

3

55

Notes

For mVs select probe type ‘Voltage’.

For mAs select probe type ‘Current’.

For mWs select mathematical function
x, and probe type ‘Voltage’ for one
channel and ‘Current’ for the other
channel.

Using Cursors on a Mathematical Result (+ - x)
Waveform

Cursor measurements on, for examle, a AxB waveform
give a reading in Watts if input A measures (milli)Volts and
input B measures (milli)Amperes.

For other cursor measurements on, for example, a A+B,
A-B or AxB waveform no reading will be available if the
input A and input B measurement unit are different.

Using Cursors on Spectrum Measurements

To do a cursor measurent on a spectrum, do the following:

1

CURSOR

From Spectrum measurement
display the cursor key label.

2

Move the cursor and observe the
readings at the top of the screen.