Using cursors on spectrum measurements – Fluke Biomedical 190M User Manual

Page 65

Advertising
background image

Using Replay, Zoom and Cursors

Making Cursor Measurements

3

55

Notes

−

For mVs select probe type ‘Voltage’.

−

For mAs select probe type ‘Current’.

−

For mWs select mathematical function
x, and probe type ‘Voltage’ for one
channel and ‘Current’ for the other
channel.

Using Cursors on a Mathematical Result (+ - x)
Waveform

Cursor measurements on, for examle, a AxB waveform
give a reading in Watts if input A measures (milli)Volts and
input B measures (milli)Amperes.

For other cursor measurements on, for example, a A+B,
A-B or AxB waveform no reading will be available if the
input A and input B measurement unit are different.

Using Cursors on Spectrum Measurements

To do a cursor measurent on a spectrum, do the following:

1

CURSOR

From Spectrum measurement
display the cursor key label.

2

Move the cursor and observe the
readings at the top of the screen.