Using The Recorder Functions

Recording Scope Waveforms In Deep Memory (Scope Record)

2

47

Using Triggering to Start or Stop Scope Record

To record an electrical event that causes a fault, it might
be useful to start or stop recording on a trigger signal:

Start on trigger to start recording; recording stops when
the deep memory is full

Stop on trigger to stop recording.

Stop when untriggered to continue recording as long as a
next trigger comes within 1 division in view all mode.

For the models 190M-4 the signal on the BNC input that
has been selected as trigger source must cause the
trigger.

For the models 190M-2 the signal applied to the banana
jack inputs (

EXT TRIGGER

(in)

). signal must cause the

trigger. The trigger source is automatically set to

Ext.

(external).

To set up the test tool, continue from step 3 of the
previous section:

4

Apply the signal to be recorded to the BNC
input(s).

5

F1

Stop recording to unlock the
OPTIONS… softkey

6

F2

Open the

RECORDER OPTIONS

menu.

7

ENTER

Jump to the Mode: field, select
on Trigger… (models 190M-4) or
on Ext. (models 190M-2) to open
the

START SINGLE SWEEP ON

TRIGGERING

or the

START SINGLE

SWEEP ON EXT

.

menu.

8

ENTER

Select one of the Conditions:
and accept the selection.

For external triggering (190M-2) continue at step 9.

9

ENTER

Select the desired trigger slope
(Slope:) and jump to Level: