Working with noisy waveforms – Fluke Biomedical 190M User Manual

Page 37

Advertising
background image

Using the Scope and Meter

Acquiring Waveforms

1

27

4

F4

Open the

INPUT A

menu.

5

ENTER

Select and accept Variable.

6

F4

Exit the menu.

At the bottom left of the screen the text A Var is displayed.
Selecting Variable will turn off cursors and automatic input
ranging.

7

mV

V

RANGE

Press mV to increase the
sensitivity, press V to decrease
the sensitivity.

Note

Variable input sensitivity is not available in the

Mathematics functions (+ - x and Spectrum).

Working with Noisy Waveforms

To suppress high frequency noise on waveforms, you can
limit the working bandwidth to 20 kHz or 20 MHz. This
function smoothes the displayed waveform. For the same
reason, it improves triggering on the waveform.
To choose HF reject on for example input A, do the
following:

1

A

Display the

INPUT A

key labels.

2

F4

Open the

INPUT A

menu.

3

ENTER

Jump to Bandwidth: and select
20kHz (HF reject) to accept the
bandwidth limitation.

Tip

To suppress noise without loss of bandwidth, use
the average function or turn off Display Glitches.