190M Series Medical ScopeMeter
Users Manual

34

4

ENTER

Select on A. Observe that the
highlight jumps to the present
measurement.

5

ENTER

Select the A dc… measurement.

6

ENTER

Select a current probe sensitivity
that matches the connected
current probe (see Adjusting the
Probe Type Settings on page 16.)

You will see a screen like in Figure 12.

Figure 12. Meter Screen

Making Relative Meter Measurements

A relative measurement displays the present
measurement result relative to a defined reference value.
The following example shows how to perform a relative
voltage measurement. First obtain a reference value:

1

METER

Display the

METER

key labels.

2

Measure a voltage to be used as
reference value.

3

F2

Set

RELATIVE

to

ON

. (

ON

is

highlighted.) This stores the
reference value as reference for
subsequent measurements.
Observe the

ADJUST REFERENCE

soft key (F3) that enables you to
adjust the reference value (see
step 5 below).

4

Measure the voltage to be
compared to the reference.

Advertising