190M Series Medical ScopeMeter
Users Manual

56

Making Rise Time Measurements

To measure rise time, do the following:

1

CURSOR

From scope mode, display the
cursor key labels.

2

F1

Press to highlight .

3

F4

For multiple traces select the
required trace A, B, C, D or M (if
a math function is active).

4

F3

Select MANUAL or AUTO (this
automatically does steps 5 to 7).

5

Move the upper cursor to 100% of
the trace height. A marker is
shown at 90%.

6

F2

Highlight the other cursor.

7

Move the lower cursor to 0% of
the trace height. A marker is
shown at 10%.

The reading shows the risetime from 10%-90% of the
trace amplitude.

Figure 27. Risetime Measurement